ISO Smart Plus
The new generation of ISO Smart dust and gas sampler has been designed and developed in accordance with international standards for sampling stack emissions.
The sampler is equipped with five thermucuple inputs for flue gas temperature, heated sampling probe and filter holder temperature measurement and control. Measurement of the gas temperature that passes through the impingers also takes place with two additional temperature sensors and inputs.
Flue gas velocity measurement takes place with a pitot tube. The measured flue gas velocity is evaluated by the microprocessor and provides isokinetic sampling. Total gas volume of the sampler is measured by the dry gas meter. All measured temperatures are stored in the device memory. All of this process is automatic and controlled by the device's microprocessor.
The sampler has built-in water trap and sensor that detects the presence of liquids passing through the inlet. In the case of liquids pass througe the inlet by accident, this sensor automatically stop the pump and gives an alarm. The user can remove the water from the water trap from drain outlet. After removing the water the samplear start to operate automatically. This important feature of the device prevents damage to the internal parts of the device.
Diffrential Pressure
dP Pitot | -100 ÷ 1000 Pa (-10÷100 mmH₂O) |
Accuracy | Better than 1% of measure ± 2Pa |
Resolution | 0.1 Pa (0.01 mmH2O) |
ABS, Pressure(Static or Barometric)
Range | 0 ÷ 105 kPa (1050 mBar) |
Accuracy | Better than 1% of measure ± 0.01 kPa |
Resolution | 0.01 kPa (0.1 mBar) |
Tempreature: (Programmed Curve Type "K" as per ITS 1990)
Range | 0 ÷ 1200 °C |
Accuracy | 1% of measure (± 0.2 °C) |
Resolution | 0.1 °C |
DGM Temperature (PT100 Sensor)
Range | -20 ÷ 80 °C |
Accuracy | 1% (± 0.2) |
Resolution | 0.1 °C |
Sampled Volume
Gas Meter Class | Gas Meter Class Dry Gas Meter, class G4, certified 2004/22/CE, in compliance to EN 1359 |
Flowrate range | 0.04 m3/h ÷ 6.0 m3/h |
Accuracy | 2% |
Encoder resolution | 0.02 liters |
Sampling Flowrate
Device | Flowmeter in compliance with UNI EN ISO 5167-2 |
Range | 5 ÷ 60 l/min |
Accuracy | Better than ± 1% |
Resolution | 0.01 l/min |
Flowrate operative range | 5 ÷ 60 l/min |
Sampled gas conditions | Dehydrated, max temperature 45°C |
Gas inlet | With protection filter for particulate and liquids |
Connection | With quick connectors |
Operation conditions | - 20 ÷ 40°C 95% UR |
Stock conditions | - 10 ÷ 50°C 95% UR |
Display | 5.0” Graphic Color |
Data Port | USB 2.0 |
Printer Port | Bluetooth |
Internal Memory | 16GB |
Power supply | 220 Vac ±10% 50/60Hz |
Materials | % 100 Aluminum |
Keyboard | Polycarbonate, tactile effect keys |
Dimensions | 395 x 262 x 502 mm (W x D x H) |
Weight | 15 kg [4.5 m3/h pump] |
- Precise isokinetic condition, sampling flowrate is directly controlled with a mass flow meter
- Higher quality of the data due to automated parameters control
- It can work with all Pitot tubes and thermocouples
- It's possible to enter the information related to Pitot tubes in the internal library
- 5 independent colored thermocouple inputs
- Optional water trap chamber
- Water detection sensor
- Stopping the pump in case of possible water leakage into the device
- Large color screen
- Isokinekik and constant flow
- Optional outdoor PM sampling option
- Easy installation of interconnection socket entries