GDS Gas Dilution System

GDS Gas Dilution System

The sampler and dilution unit are in a single cabin. Flow control of both units is done with two independent mass flow meters integrated into the device. Total and dilution gas volume can measure with two independent dry gas meter. All of this process is automatic and controlled by the device's microprocessor.

The two streams are regulated independently, so for different stack condition the dilution ratio can be changed indepandantly. The temperatures inside both dry gas meters reads with two tempreture sensor and stores in the memory of the sampler.

The dilution and total sample inlets are equipped with two easily replaceable cartridges. In addition, the user can easily replace the cartridge silica gel and charcoal. Unlike traditional manual sampler, it is fully automated. Dilution and total flow rates can be adjusted digitally. The sampler automatically controls the set flow rates and adjusts the flow rates. The sampler also directly reads gas volumes and temperatures from the dry gas meter. Thus, it is not necessary to note the first and last readings and temperatures from the dry gas meter.

Total Sampled Volume

Gas Meter Class Gas Meter Class Dry Gas Meter, class G4, certified 2004/22/CE, in compliance to EN 1359
Flowrate range 0.016 m3/h ÷ 2.5 m3/h
Accuracy 2%
Encoder resolution 0.02 liters

Dilution Sampling Flowrate

Device Flowmeter in compliance with UNI EN ISO 5167-2
Range 0.1 ÷ 1.2 l/min
Accuracy Better than ± 1%
Resolution 0.01 l/min

Total Sampling Flowrate

Device Flowmeter in compliance with UNI EN ISO 5167-2
Range 0.1 ÷ 1.2 l/min
Accuracy Better than ± 1%
Resolution 0.01 l/min

ABS, Pressure(Static or Barometric)

Range 0 ÷ 105 kPa (1050 mBar)
Accuracy Better than 1% of measure ± 0.01 kPa
Resolution 0.01 kPa (0.1 mBar)

Tempreature: (Programmed Curve Type "K" as per ITS 1990)

Range 0 ÷ 1200 °C
Accuracy 1% of measure (± 0.2 °C)
Resolution 0.1 °C

DGM Temperature (PT100 Sensor)

Range -20 ÷ 80 °C
Accuracy 1% (± 0.2)
Resolution 0.1 °C
Flowrate operative range 0.1 ÷ 1.2 l/min
Sampled gas conditions Dehydrated, max temperature 45°C
Total Gas inlet With protection filter and silicagel cartridge for particulate and liquids
Connection With quick connectors
Operation conditions - 20 ÷ 40°C 95% UR
Stock conditions - 10 ÷ 50°C 95% UR
Display 5.0” Graphic Color
Keyboard Polycarbonate, tactile effect keys
Data Port USB 2.0
Printer Port Bluetooth
Internal Memory 16GB
Power supply 220 Vac ±10% 50/60Hz
Materials % 100 Aluminum
Dimensions 395 x 262 x 502 mm (W x D x H)
Weight 10 kg
  • Precise sampling flowrate is directly controlled with a mass flow meter
  • Higher quality of the data due to automated parameters control
  • It's possible to enter and set the dilution and totale sample flowrate
  • Total sampel gas inlet protected by removeble silicagel cartridge
  • Removeble charcoal cartridge for zero gas outlet
  • 3 independent colored thermocouple inputs
  • Large color screen
  • Easy installation of interconnection socket entries