ISO Smart Plus

ISO Smart Plus

The new generation of ISO Smart dust and gas sampler has been designed and developed in accordance with international standards for sampling stack emissions.
The sampler is equipped with five thermucuple inputs for flue gas temperature, heated sampling probe and filter holder temperature measurement and control. Measurement of the gas temperature that passes through the impingers also takes place with two additional temperature sensors and inputs.
Flue gas velocity measurement takes place with a pitot tube. The measured flue gas velocity is evaluated by the microprocessor and provides isokinetic sampling. Total gas volume of the sampler is measured by the dry gas meter. All measured temperatures are stored in the device memory. All of this process is automatic and controlled by the device's microprocessor.
The sampler has built-in water trap and sensor that detects the presence of liquids passing through the inlet. In the case of liquids pass througe the inlet by accident, this sensor automatically stop the pump and gives an alarm. The user can remove the water from the water trap from drain outlet. After removing the water the samplear start to operate automatically. This important feature of the device prevents damage to the internal parts of the device.

Diffrential Pressure

dP Pitot -100 ÷ 1000 Pa (-10÷100 mmH₂O)
Accuracy Better than 1% of measure ± 2Pa
Resolution 0.1 Pa (0.01 mmH2O)

ABS, Pressure(Static or Barometric)

Range 0 ÷ 105 kPa (1050 mBar)
Accuracy Better than 1% of measure ± 0.01 kPa
Resolution 0.01 kPa (0.1 mBar)

Tempreature: (Programmed Curve Type "K" as per ITS 1990)

Range 0 ÷ 1200 °C
Accuracy 1% of measure (± 0.2 °C)
Resolution 0.1 °C

DGM Temperature (PT100 Sensor)

Range -20 ÷ 80 °C
Accuracy 1% (± 0.2)
Resolution 0.1 °C

Sampled Volume

Gas Meter Class Gas Meter Class Dry Gas Meter, class G4, certified 2004/22/CE, in compliance to EN 1359
Flowrate range 0.04 m3/h ÷ 6.0 m3/h
Accuracy 2%
Encoder resolution 0.02 liters

Sampling Flowrate

Device Flowmeter in compliance with UNI EN ISO 5167-2
Range 5 ÷ 60 l/min
Accuracy Better than ± 1%
Resolution 0.01 l/min
Flowrate operative range 5 ÷ 60 l/min
Sampled gas conditions Dehydrated, max temperature 45°C
Gas inlet With protection filter for particulate and liquids
Connection With quick connectors
Operation conditions - 20 ÷ 40°C 95% UR
Stock conditions - 10 ÷ 50°C 95% UR
Display 5.0” Graphic Color
Data Port USB 2.0
Printer Port Bluetooth
Internal Memory 16GB
Power supply 220 Vac ±10% 50/60Hz
Materials % 100 Aluminum
Keyboard Polycarbonate, tactile effect keys
Dimensions 395 x 262 x 502 mm (W x D x H)
Weight 15 kg [4.5 m3/h pump]
  • Precise isokinetic condition, sampling flowrate is directly controlled with a mass flow meter
  • Higher quality of the data due to automated parameters control
  • It can work with all Pitot tubes and thermocouples
  • It's possible to enter the information related to Pitot tubes in the internal library
  • 5 independent colored thermocouple inputs
  • Optional water trap chamber
  • Water detection sensor
  • Stopping the pump in case of possible water leakage into the device
  • Large color screen
  • Isokinekik and constant flow
  • Optional outdoor PM sampling option
  • Easy installation of interconnection socket entries